AED500 Automatic External Defibrillator
Defibrillator
Category
|
Specification |
Output | Biphasic Truncated exponential |
Energy Sequence(child ) | 50J ,50J ,50J
|
Energy Sequence(Adult)
| 150J ,200J ,200J
|
Charge Time
| 6 sec. to 150J 8 sec. to 200 J(New battery)
|
Analysis Time
| 6 sec.
|
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery.
| Less than 10 seconds
|
The maximum time from the in initiation on of rhythm analysis to readiness for discharge after 6 shocks.
| Less than 10 seconds
|
The maximum time from initially switching power onto readiness for discharge.
| Less than 18 seconds
|
Visual Prompts
| LED prompts;Automatic Self-test
|
Controls
| Four buttons : On/Off, Shock , Adult/child , CPR
|
Output Energy Accuracy
| ±10% into any impedance from 25 to 1 75Q |
The Maximum Voltage
| 1400±50V
|
Output disabled when the patient impedance is outsidelimits
|
20Ω to 200Ω |
Battery
| |
Part No
| CR123A-4x2
|
Non-rechargeable
| 2. 8 Ah
|
Capacity
| 120 discharges at 200 Joules
|
Shelf Life (25°C +15C)
| 5 years standby (after installation ) |
Physical
| |
Dimensions
| 280 x 209x 80 mm
|
Weight
| 2.5 kg
|
Operating Temperature
| 0 ℃to 40 ℃
|
Operating Humidity
| Relative humidity between 30% and 95%(non-condensing)
|
Storage Temperature(without battery)
| -20℃ to 55 ℃
|
Storage Humidity(without battery)
| Up to 93%(non-condensing) |
Packing:Single package size:40*30*17.8 cm 6 unit per carton,package size:62.5°42.5 *56.5cm,total gross weight 21 KG
AED500 Automatic External Defibrillator
Defibrillator
Category
|
Specification |
Output | Biphasic Truncated exponential |
Energy Sequence(child ) | 50J ,50J ,50J
|
Energy Sequence(Adult)
| 150J ,200J ,200J
|
Charge Time
| 6 sec. to 150J 8 sec. to 200 J(New battery)
|
Analysis Time
| 6 sec.
|
The maximum time from the initiation of rhythm analysis to readiness for discharge with a new battery.
| Less than 10 seconds
|
The maximum time from the in initiation on of rhythm analysis to readiness for discharge after 6 shocks.
| Less than 10 seconds
|
The maximum time from initially switching power onto readiness for discharge.
| Less than 18 seconds
|
Visual Prompts
| LED prompts;Automatic Self-test
|
Controls
| Four buttons : On/Off, Shock , Adult/child , CPR
|
Output Energy Accuracy
| ±10% into any impedance from 25 to 1 75Q |
The Maximum Voltage
| 1400±50V
|
Output disabled when the patient impedance is outsidelimits
|
20Ω to 200Ω |
Battery
| |
Part No
| CR123A-4x2
|
Non-rechargeable
| 2. 8 Ah
|
Capacity
| 120 discharges at 200 Joules
|
Shelf Life (25°C +15C)
| 5 years standby (after installation ) |
Physical
| |
Dimensions
| 280 x 209x 80 mm
|
Weight
| 2.5 kg
|
Operating Temperature
| 0 ℃to 40 ℃
|
Operating Humidity
| Relative humidity between 30% and 95%(non-condensing)
|
Storage Temperature(without battery)
| -20℃ to 55 ℃
|
Storage Humidity(without battery)
| Up to 93%(non-condensing) |
Packing:Single package size:40*30*17.8 cm 6 unit per carton,package size:62.5°42.5 *56.5cm,total gross weight 21 KG